Basler Light Dark Field-50OD-x#

Basler Dark Field Lights can be used in various applications for dark field lighting.

Basler Light Dark Field

Key Features#

  • Integrated Light Control (ILC): Direct light control via Basler camera, eliminating the need for an external light controller.
  • Integrated LED driver with two light control modes:
  • Overdrive option in both light control modes:
    Increased light output without requiring additional hardware (see Overdriving Light Devices)
  • Automatic light detection: Camera automatically identifies the connected Basler Light and applies appropriate operating and protection settings.
  • GenICam-compliant control via Basler cameras: Full control of brightness, strobe and continuous-light operation directly from a Basler camera using any GenICam-compatible software, such as the pylon Software Suite
  • Integrated protection mechanisms: Over-triggering and thermal overload protection ensure reliable operation and protect the light throughout its service life.
  • Support for external controllers: Can also be operated in External Control Mode via any external light controller.

Basler Lights Before and After Revision 05#

Info

To determine whether a Basler Light has Integrated Light Control (ILC), check the revision number on the light. The last two digits following the order number indicate the revision. In the example 2200000841-05, the revision is 05.

  • Revision 05 or later: The light has ILC and an integrated controller.
  • Revision below 05: The light doesn't have ILC. It requires an external light controller.

Documentation for Basler Lights without ILC is no longer included in the current Basler Product Documentation. For information about these lights, download the offline documentation of an earlier version by clicking here.

Changes to the Pin Assignment#

Info

Basler Lights with revision 05 or later use a different pin assignment than Basler Lights with a revision below 05.

Adjust the pin assignment when both of the following are true:

  • You replace a Basler Light with a revision below 05 with a Basler Light with revision 05 or later.
  • You operate the light with an external light controller, such as the Basler SLP Strobe Controller or a third-party controller.

Verify that you are using the correct pin assignment when connecting a light. If you keep the original pin assignment, the new light doesn't turn on.

For information about the pin assignment, see Operating Basler Lights with Internal Controller or Operating Basler Lights with External Controller.

General Specifications#

Basler Light Dark Field-50OD-Red Basler Light Dark Field-50OD-White Basler Light Dark Field-50OD-Blue Basler Light Dark Field-50OD-Infrared
Order Number, IP54 Lights 2200000849 2200000745 2200000818 2200000880
Peak Wavelength 625 nm - 465 nm 850 nm
Color Temperature - 5000 K - -
LED Color Red White Blue Infrared
Luminous Area (outer diameter) 50 mm 50 mm 50 mm 50 mm
Recommended Light Working Distance 1–30 mm 1–30 mm 1–30 mm 1–30 mm
Risk Class (according to IEC 62471) Exempt Exempt Exempt Exempt
Basler Light Dark Field-50OD-x
Conformity CE (includes RoHS), FCC, IP54, UKCA
Certificates for your light model
For more information, see the Compliance section of the Basler website.

LED Specifications#

Dark Field Light Series 50OD-x
Luminous Power of
White LEDs (lumen)
725
Radiant Power of
Red LEDs (mW)
2130
Radiant Power of
Blue LEDs (mW)
TBD
Radiant Power of
IR LEDs (mW)
1030

Electrical Specifications#

White,
5000 K
Red,
625 nm
Blue,
465 nm
IR,
850 nm
Required Voltage 24 VDC
± 5 %
24 VDC
± 5 %
24 VDC
± 5 %
24 VDC
± 5 %
Required Power for
Continuous Light
6.0 W 4.3 W 5.9 W 4.9 W
Max. LED Current for
Continuous Light
(100 % Brightness)
300 mA 300 mA 300 mA 450 mA
Overdrive Operation Ex. 1:
Max. LED Current, Duty Cycle (%),
Max. Pulse Width
450 mA,
60 %,
200 ms
450 mA,
60 %,
200 ms
450 mA,
60 %,
200 ms
675 mA,
40 %,
10 ms
Overdrive Operation Ex. 2:
Max. LED Current, Duty Cycle (%),
Max. Pulse Width
600 mA,
40 %,
100 ms
600 mA,
40 %,
100 ms
600 mA,
40 %,
100 ms
900 mA,
20 %,
5 ms
Overdrive Operation Ex. 3:
Max. LED Current, Duty Cycle (%),
Max. Pulse Width
900 mA,
20 %,
50 ms
900 mA,
20 %,
50 ms
900 mA,
20 %,
50 ms
1350 mA,
10 %,
1 ms
Overdrive Operation Ex. 4:
Max. LED Current, Duty Cycle (%),
Max. Pulse Width
1200 mA,
10 %,
10 ms
1200 mA,
10 %,
10 ms
1200 mA,
10 %,
10 ms
1800 mA,
5 %,
200 μs
Recommended Min. Flash Time in
Overdrive Operation (see info below)
100 μs 100 μs 100 μs 100 μs

Info

  • The examples 1 to 4 in the table above are reference limit values only.
    Lower values for LED current, duty cycle and maximum pulse width are also possible.
  • About the recommended minimum flash time in overdrive operation: The latency between the trigger signal and the LED light going on is about maximum 30 μs. Therefore, the minimum recommended flash time in the table is 100 μs.

Mechanical Specifications#

Dark Field Light Series 50OD-x
Dimensions
(H x W x D)
130 x 141 x 12 mm
Weight 360 g
Material Black anodized aluminum housing with internal PMMA light guide
Connector M8 socket, 4-pin male connector

Dark Field Light Dimensions#

Basler Light Darkfield 50OD

Connector Pin Numbering#

Connector Pin Numbering Basler Light

The current Basler Light uses the M8 4-pin connector model from CONEC (article no. 42-01033).

Info

The pin assignment of the light connector depends on whether the light is operated using its integrated controller (internal control) or an external controller. For details, refer to the relevant tables in the topics of the Controlling Basler Lights section:

Environmental Requirements#

Temperature During Operation 10–30 °C (50–86 °F) / 45 °C (113 °F)a
Humidity During Operation 30–70 %

  1. For steady light operation without additional heat dissipation measures, a maximum of 30 °C (86 °F) is recommended. For a maximum of 45 °C (113 °F), heat dissipation measures are mandatory.
    For flash light operation with a maximum 10 % duty cycle, a maximum of 45 °C (113 °F) is also permissible.

Performance Charts#

Peak Illuminance versus Light Working Distance (LWD)#

Peak Illuminance versus Light Working Distance (LWD)

Info

Diagram is based on the following reference model: BL Dark Field-50OD-White.

Spectrum#

Spectrum

Uniformity#

Uniformity

Info

The values in the graph above are measured in mm.

Uniformity Intensity